Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Published September 21, 2022 · Recall initiated: July 19, 2022
Summary
FDA records medical device Class II recall Z-1715-2022, published September 21, 2022 and initiated July 19, 2022. Recalling firm: Cook Incorporated; Status: Ongoing; Reported quantity: 865 units; Lot: UDI-DI: (01)00827002561762- Lot Numbers: 14112845, NS14062106, NS14080814, NS14089120, NS14118843, NS14139358, NS14142006, NS14144747, NS14148490, NS14149799, NS14153797, NS14172915. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Incorporated
- Distributor
- Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
- Lot
- UDI-DI: (01)00827002561762- Lot Numbers: 14112845, NS14062106, NS14080814, NS14089120, NS14118843, NS14139358, NS14142006, NS14144747, NS14148490, NS14149799, NS14153797, NS14172915
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAWilliams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, andOctober 21, 2026
- FDACentimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographicMay 13, 2026
- FDAAurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters areMay 13, 2026
- FDABeacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20May 13, 2026
- FDARoadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516. ForSeptember 21, 2022
- FDAMicroClave Clear Neutral Connector; MicroClave Clear Connector; MicroClave Clear Neutral Connector, ApproxSeptember 21, 2022
Questions about this recall
Was Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter positioning and exchange in diagnostic and interventional procedures. recalled?
Yes. FDA published recall Z-1715-2022 on September 21, 2022. Reported hazard: Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
What should I do if I have this product?
Voluntary: Firm initiated · Letter