Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively.
Published June 11, 2014 · Recall initiated: May 9, 2014
Summary
FDA records medical device Class II recall Z-1718-2014, published June 11, 2014 and initiated May 9, 2014. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 2; Lot: lot 380652-20. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
What to do
Voluntary: Firm initiated · Other
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- Medical facilities in TX and FL
- Lot
- lot 380652-20
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDASiemens Ysio Systems Product Usage: The Ysio (New RAD-FAMILY) systems are radiographic systems used inJune 11, 2014
- FDAMACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patientJune 11, 2014
Questions about this recall
Was Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively. recalled?
Yes. FDA published recall Z-1718-2014 on June 11, 2014. Reported hazard: Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
What should I do if I have this product?
Voluntary: Firm initiated · Other