Siemens SOMATOM Emotion 16, Emotion 6, and SOMATOM Spirit Computed Tomography Systems Product Usage: To produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

Published June 11, 2014 · Recall initiated: March 21, 2014

Summary

FDA records medical device Class II recall Z-1721-2014, published June 11, 2014 and initiated March 21, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 345; Lot: Material numbers: 10165977, 10045692 , and 10165888 with serial numbers: 32187 31158 31215 31222 31313 31354 32118 32120 32123 32125 32126 32127 32129 32139 29161 27229 29185 27415 27212 29364 29186 29165 29233 29182 27235 29338 29176 29376 27232 29181 27369 29295 27230 29198 27244 29286 29126 27338 29302 27231 27334 29394 29264 32141 32156 32159. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution in the states of IA, TX, FL, NE, NY, OH, NC, KS, IL, CA, OK including PR.
Lot
Material numbers: 10165977, 10045692 , and 10165888 with serial numbers: 32187 31158 31215 31222 31313 31354 32118 32120 32123 32125 32126 32127 32129 32139 29161 27229 29185 27415 27212 29364 29186 29165 29233 29182 27235 29338 29176 29376 27232 29181 27369 29295 27230 29198 27244 29286 29126 27338 29302 27231 27334 29394 29264 32141 32156 32159
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens SOMATOM Emotion 16, Emotion 6, and SOMATOM Spirit Computed Tomography Systems Product Usage: To produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. recalled?

Yes. FDA published recall Z-1721-2014 on June 11, 2014. Reported hazard: The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter