FDAFDA · Food and drugs·Recall number Z-1722-2014

Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners)

Published June 11, 2014 · Recall initiated: April 1, 2014

Summary

FDA records medical device Class II recall Z-1722-2014, published June 11, 2014 and initiated April 1, 2014. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 6, 839 units; Lot: Brilliance CT 6: M/N: 728256; S/N: 3001, 3002, 3005, 3007, 3014, 3016, 3019-3021, 3025, 3026, 3028, 3032-3036, 3045, 3046, 3049, 3051, 3052, 3057, 3058, 3060, 3062, 3064-3066, 3068, 3071, 3073, 3074, 3076-3081, 3083, 3084, 3090, 3093-3095, 3097, 3101, 3102, 3104, 3106-3108, 3112-3115, 3118-3121, 3124, 3126, 3128, 3129, 3131, 3132, 3134, 3136, 3137, 3139-3141, 3143-3145, 3147, 3148, 3151-3158, 3162, 3165-3169, 3171, 3173-3177, 3180, 3181, 3184-3190, 3192, 3193, 3196, 3200, 3201, 3203, 3204, 3207, 3208, 3210-3218, 3220, 3221, 3225, 3226, 3228, 3229, 3232, 3235, 3237, 3240, 3241, 3243-3245, 3248, 3249, 3253-3257, 3260-3264, 3266-3272, 3275-3282, 3284, 3285, 3287, 3290, 3292, 3296, 3299, 3301, 3303-3305, 3307-3311, 3313, 3315, 3317-3323, 3325-3328, 3332-3334, 3336, 3338, 3339, 3341, 3343-3345, 3347-3349, 3351, 3352, 3354, 3357-3364, 3366-3368, 3370-3374, 3377-3384, 3386, 3388-3391, 3394, 3396-3400, 3402-3404, 3408, 3409, 3412, 3413, 3415-3418, 3423-3426, 3428-3432, 3434-3437, 3439, 3443, 3444, 3447, 3449, 3451-3453, 3455, 3457, 3459, 3460, 3461, 3463, 3465, 3467-3475, 3477-3483, 3485, 3486, 3488, 3489, 3491-3506, 3508-3511, 3513, 3515, 3517-3519, 3522-3524, 3526, 3527, 3532-3535, 3537-3549, 3551-3553, 3556-3559, 3562, 3564, 3566, 3568, 3570, 3571, 3573-3575, 3577, 3578, 3580, 3582-3588, 3590-3595, 3597-3608, 3611-3616, 3618-3622, 3625-3632, 3634-3637, 3639-3646, 3648, 3649, 3651-3661, 3663-3665, 3667-3670, 3672, 3674-3677, 3679-3693, 3695-3698, 3703-3712, 3714, 3715, 3717-3720, 3722, 6039, 6040, 30003, 30005, 30007-30009, 30011-30018, 30021, 30022, 30024, 30026-30031, 30033-30040, 30042, 30043, 30045-30048, 30050, 30052-30054, 30056-30062, 30064-30066, 30068, 30069, 30072, 30073, 30075, 30077-30080, 30082-30090, 30092, 30094, 30100, 30101, 30104-30108, 30111, 30113, 30117-30119, 30121-30131 & 300043. Brilliance CT 10: M/N: 728251; S/N: 3008, 3098, 4001, 4003, 4005, 4006, 4009-4012, 4014, 4016, 4020, 4021, 4023, 4026, 4027, 4029-4031, 4035, 4040, 4041, 4044, 4047-4050, 4052-4054, 4058, 4059, 4061, 4063-4067, 30099, 40006-40012, 40014-40018, 40021, 40023, 40024, 40028, 40030 & 40032-40035. Brilliance CT 16: M/N: 728246; S/N: 3011, 3013, 3018, 3027, 3031, 3038, 3041, 3043, 3070, 3072, 3082, 3085, 3117, 3127, 3149, 3160, 3205, 3216, 3227, 3293-3295, 3302, 3306, 3324, 3336, 3376, 3393, 3466, 3490, 3514, 3521, 3563, 3565, 3624, 3647, 3673, 3721, 4004, 4017, 4018, 4019, 4024, 4025, 4036, 4043, 4055, 5001, 5004-5008, 5010-5021, 5023-5026, 5028-5032, 5034-5038, 5040-5042, 5044-5046, 5049-5055, 5058-5062, 5064, 5066-5070, 5072-5076, 5078, 5081-5086, 5088-5095, 5097-5099, 5101-5105, 5107, 5108, 5110-5118, 5120-5128, 5130-5141, 5143, 5145-5152, 5155, 5156, 5158, 5160-5164, 5166-5170, 5173-5181, 5183-5202, 5204, 5207-5209, 5211-5219, 5221-5226, 5228-5240, 5242, 5243, 5248, 5249, 5252-5254, 5258, 5264-5267, 5269-5275, 5278, 5279, 5281-5283, 5285-5297, 5301-5309, 5311-5313, 5315-5320, 5322, 5323, 5325-5332, 5334, 5336, 5338-5344, 5346-5351, 5353, 5355, 5357, 5359-5362, 5365-5369, 5371, 5372, 5375-5380, 5382, 5384, 5386-5389, 5394, 5396-5400, 5402-5410, 5414-5429, 5432-5435, 5437-5443, 5451, 5452, 5456, 5460, 5468, 5469, 5485, 5489, 5515, 5522, 5531, 5535, 5537, 5543, 5552, 5555, 5557, 5567, 5581-5584, 5586-5588, 5600, 5607, 5621, 5624, 5630, 5642, 5644, 5645, 5651, 5654, 5658, 5663, 5664, 5667, 5675, 5679, 5684, 5685, 5697, 5707, 5709, 5716, 5724, 5725, 5735, 5738, 5743-5745, 5748-5750, 5752, 5754, 5756, 5772, 5774, 5776, 5777, 5780, 5782, 5790, 5793, 5794, 5796, 5801, 5802, 5806, 5807, 5812-5815, 5823-5827, 5830-5834, 5840-5849, 5851, 5852, 5854, 5855, 5857, 5858, 5860, 5861, 5866, 5871-5873, 5880, 5881, 5882, 5884-5886, 5892, 5899, 5901, 5904, 5915, 5921-5923, 5939-5942, 5947, 5948, 5961-5963, 5966-5968, 5976, 5980, 6011, 6012, 6014, 6015, 6028, 6033, 6034, 6036-6038, 6045, 6050, 6068-6071, 6082, 6090, 6095, 6104, 6105, 6108, 6112, 6115, 6126, 6132, 6133, 6161-6165, 6173, 6180, 6221, 6240, 6241-6243, 6245-6257, 6259-6262, 6264-6273, 6. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution: US nationwide including PR and DC, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Cambodia, Cape Verde, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Greenland, Guadeloupe, Herzegovina, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, LAO People's Dem, Rep., Latvia, Lebanon, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nahar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam & Yemen.
Lot
Brilliance CT 6: M/N: 728256; S/N: 3001, 3002, 3005, 3007, 3014, 3016, 3019-3021, 3025, 3026, 3028, 3032-3036, 3045, 3046, 3049, 3051, 3052, 3057, 3058, 3060, 3062, 3064-3066, 3068, 3071, 3073, 3074, 3076-3081, 3083, 3084, 3090, 3093-3095, 3097, 3101, 3102, 3104, 3106-3108, 3112-3115, 3118-3121, 3124, 3126, 3128, 3129, 3131, 3132, 3134, 3136, 3137, 3139-3141, 3143-3145, 3147, 3148, 3151-3158, 3162, 3165-3169, 3171, 3173-3177, 3180, 3181, 3184-3190, 3192, 3193, 3196, 3200, 3201, 3203, 3204, 3207, 3208, 3210-3218, 3220, 3221, 3225, 3226, 3228, 3229, 3232, 3235, 3237, 3240, 3241, 3243-3245, 3248, 3249, 3253-3257, 3260-3264, 3266-3272, 3275-3282, 3284, 3285, 3287, 3290, 3292, 3296, 3299, 3301, 3303-3305, 3307-3311, 3313, 3315, 3317-3323, 3325-3328, 3332-3334, 3336, 3338, 3339, 3341, 3343-3345, 3347-3349, 3351, 3352, 3354, 3357-3364, 3366-3368, 3370-3374, 3377-3384, 3386, 3388-3391, 3394, 3396-3400, 3402-3404, 3408, 3409, 3412, 3413, 3415-3418, 3423-3426, 3428-3432, 3434-3437, 3439, 3443, 3444, 3447, 3449, 3451-3453, 3455, 3457, 3459, 3460, 3461, 3463, 3465, 3467-3475, 3477-3483, 3485, 3486, 3488, 3489, 3491-3506, 3508-3511, 3513, 3515, 3517-3519, 3522-3524, 3526, 3527, 3532-3535, 3537-3549, 3551-3553, 3556-3559, 3562, 3564, 3566, 3568, 3570, 3571, 3573-3575, 3577, 3578, 3580, 3582-3588, 3590-3595, 3597-3608, 3611-3616, 3618-3622, 3625-3632, 3634-3637, 3639-3646, 3648, 3649, 3651-3661, 3663-3665, 3667-3670, 3672, 3674-3677, 3679-3693, 3695-3698, 3703-3712, 3714, 3715, 3717-3720, 3722, 6039, 6040, 30003, 30005, 30007-30009, 30011-30018, 30021, 30022, 30024, 30026-30031, 30033-30040, 30042, 30043, 30045-30048, 30050, 30052-30054, 30056-30062, 30064-30066, 30068, 30069, 30072, 30073, 30075, 30077-30080, 30082-30090, 30092, 30094, 30100, 30101, 30104-30108, 30111, 30113, 30117-30119, 30121-30131 & 300043. Brilliance CT 10: M/N: 728251; S/N: 3008, 3098, 4001, 4003, 4005, 4006, 4009-4012, 4014, 4016, 4020, 4021, 4023, 4026, 4027, 4029-4031, 4035, 4040, 4041, 4044, 4047-4050, 4052-4054, 4058, 4059, 4061, 4063-4067, 30099, 40006-40012, 40014-40018, 40021, 40023, 40024, 40028, 40030 & 40032-40035. Brilliance CT 16: M/N: 728246; S/N: 3011, 3013, 3018, 3027, 3031, 3038, 3041, 3043, 3070, 3072, 3082, 3085, 3117, 3127, 3149, 3160, 3205, 3216, 3227, 3293-3295, 3302, 3306, 3324, 3336, 3376, 3393, 3466, 3490, 3514, 3521, 3563, 3565, 3624, 3647, 3673, 3721, 4004, 4017, 4018, 4019, 4024, 4025, 4036, 4043, 4055, 5001, 5004-5008, 5010-5021, 5023-5026, 5028-5032, 5034-5038, 5040-5042, 5044-5046, 5049-5055, 5058-5062, 5064, 5066-5070, 5072-5076, 5078, 5081-5086, 5088-5095, 5097-5099, 5101-5105, 5107, 5108, 5110-5118, 5120-5128, 5130-5141, 5143, 5145-5152, 5155, 5156, 5158, 5160-5164, 5166-5170, 5173-5181, 5183-5202, 5204, 5207-5209, 5211-5219, 5221-5226, 5228-5240, 5242, 5243, 5248, 5249, 5252-5254, 5258, 5264-5267, 5269-5275, 5278, 5279, 5281-5283, 5285-5297, 5301-5309, 5311-5313, 5315-5320, 5322, 5323, 5325-5332, 5334, 5336, 5338-5344, 5346-5351, 5353, 5355, 5357, 5359-5362, 5365-5369, 5371, 5372, 5375-5380, 5382, 5384, 5386-5389, 5394, 5396-5400, 5402-5410, 5414-5429, 5432-5435, 5437-5443, 5451, 5452, 5456, 5460, 5468, 5469, 5485, 5489, 5515, 5522, 5531, 5535, 5537, 5543, 5552, 5555, 5557, 5567, 5581-5584, 5586-5588, 5600, 5607, 5621, 5624, 5630, 5642, 5644, 5645, 5651, 5654, 5658, 5663, 5664, 5667, 5675, 5679, 5684, 5685, 5697, 5707, 5709, 5716, 5724, 5725, 5735, 5738, 5743-5745, 5748-5750, 5752, 5754, 5756, 5772, 5774, 5776, 5777, 5780, 5782, 5790, 5793, 5794, 5796, 5801, 5802, 5806, 5807, 5812-5815, 5823-5827, 5830-5834, 5840-5849, 5851, 5852, 5854, 5855, 5857, 5858, 5860, 5861, 5866, 5871-5873, 5880, 5881, 5882, 5884-5886, 5892, 5899, 5901, 5904, 5915, 5921-5923, 5939-5942, 5947, 5948, 5961-5963, 5966-5968, 5976, 5980, 6011, 6012, 6014, 6015, 6028, 6033, 6034, 6036-6038, 6045, 6050, 6068-6071, 6082, 6090, 6095, 6104, 6105, 6108, 6112, 6115, 6126, 6132, 6133, 6161-6165, 6173, 6180, 6221, 6240, 6241-6243, 6245-6257, 6259-6262, 6264-6273, 6
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners) recalled?

Yes. FDA published recall Z-1722-2014 on June 11, 2014. Reported hazard: A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.

What should I do if I have this product?

Voluntary: Firm initiated · Letter