BEQ-TOP 27610 ECC PACK 3/8, Custom Tubing Kit, Catalog No. 701065622
Published May 16, 2018 · Recall initiated: October 19, 2017
Summary
FDA records medical device Class II recall Z-1721-2018, published May 16, 2018 and initiated October 19, 2017. Recalling firm: Datascope Corporation; Status: Terminated; Reported quantity: 9; Lot: Lot 3000058734, 3000063711,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Datascope Corporation
- Distributor
- Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
- Lot
- Lot 3000058734, 3000063711,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.November 2, 2022
- FDAMEGA 8Fr 50cc Intra-Aortic Balloon CatheterSeptember 14, 2022
- FDADatascope Intra-Aortic Ballon Catheters (IABs)-Linear 7.5Fr 34cc IAB Kit P/N: 0684-00-0479-01,September 2, 2020
- FDADatascope Intra-Aortic Ballon Catheters (IABs)-Trans-Ray 7Fr 40cc IAB (Japan) - Product Usage: Indications ForSeptember 2, 2020
- FDABEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179May 16, 2018
- FDABEQ-TOP 22300, Custom Tubing Kit, Catalog No. 701049504May 16, 2018
Questions about this recall
Was BEQ-TOP 27610 ECC PACK 3/8, Custom Tubing Kit, Catalog No. 701065622 recalled?
Yes. FDA published recall Z-1721-2018 on May 16, 2018. Reported hazard: The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
What should I do if I have this product?
Voluntary: Firm initiated · Letter