Canon Aquilion Prime, Model TSX-303A

Published April 22, 2020 · Recall initiated: March 25, 2019

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Canon Medical System, USA, INC.
Distributor
Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Lot
Canon Aquilion PRIME, TSX-303A - Serial numbers 1CB1892050, FLC1882279, FLC1892282, FLC1882278, FLC18Y2301, FLC1892288, FOC1892284, FOC1872277, FOC18X2289, and FOC18Y2296.
Risk classification
Class II
Area (FDA)
device
Country
United States