FDAFDA · Food and drugs·Recall number Z-1724-2020

Canon Aquilion Prime SP, Model TSX-303B

Published April 22, 2020 · Recall initiated: March 25, 2019

Summary

FDA records medical device Class II recall Z-1724-2020, published April 22, 2020 and initiated March 25, 2019. Recalling firm: Canon Medical System, USA, INC.; Status: Ongoing; Reported quantity: 10 devices; Lot: Canon Aquilion Prime SP, Multislice Helical CT Scanner, Model TSX-303B - Serial numbers 1LB1882092, 1LB1882088, 1LB18X2111, 1LB1882086, 1LB18X2129, 1LB18X2109, 1LB1882087, 1LB1882090, 1LB1892093, and 1OB1892102. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Canon Medical System, USA, INC.
Distributor
Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Lot
Canon Aquilion Prime SP, Multislice Helical CT Scanner, Model TSX-303B - Serial numbers 1LB1882092, 1LB1882088, 1LB18X2111, 1LB1882086, 1LB18X2129, 1LB18X2109, 1LB1882087, 1LB1882090, 1LB1892093, and 1OB1892102.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Canon Aquilion Prime SP, Model TSX-303B recalled?

Yes. FDA published recall Z-1724-2020 on April 22, 2020. Reported hazard: Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

What should I do if I have this product?

Voluntary: Firm initiated · Letter