FDAFDA · Food and drugs·Recall number Z-1729-2014
HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.
Published June 18, 2014 · Recall initiated: April 24, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Cenorin, LLC
- Distributor
- Distribution USA Nationwide
- Lot
- 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States