HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Published July 16, 2014 · Recall initiated: January 24, 2013
Summary
FDA records medical device Class II recall Z-1995-2014, published July 16, 2014 and initiated January 24, 2013. Recalling firm: Cenorin, LLC; Status: Terminated; Reported quantity: 25 HLD systems; Lot: 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cenorin, LLC
- Distributor
- Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
- Lot
- 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 ForcedAugust 23, 2017
- FDAThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010 ForcedAugust 23, 2017
- FDAPatient Side Cart Battery Box used in conjunction with the da Vinci A surgical system. Intuitive Surgical Inc.July 16, 2014
- FDAHenora REF: 70-4026-09V1 300psi I.V. Catheter Extension Set Product Usage: Catheter extension set for highJuly 16, 2014
- FDAAmerigel Wound Dressing 55 gallons/drumJuly 16, 2014
- FDAHLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water isJune 18, 2014
Questions about this recall
Was HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer. recalled?
Yes. FDA published recall Z-1995-2014 on July 16, 2014. Reported hazard: The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit