Back to search
FDAFDA · Food and drugs·Recall number Z-1745-2013

Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone lengthening, including conditions such as congenital mandibular deficiencies or post-traumatic defects. The Multi Vector Distractor can be utilized to perform bone transport procedures as an alternative to bone grafts and free flaps.

Published July 24, 2013 · Recall initiated: May 31, 2013

Summary

FDA records medical device Class II recall Z-1745-2013, published July 24, 2013 and initiated May 31, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 121; Lot: Titanium Multi Vector Distractor Pin Holding Clamp, body, limited bone stock Part # 487.941 with lot numbers: 3769720, 3779569, 3788731, 3794434, 7806093, 7823463 and Titanium Multi Vector Distractor Pin Holding Clamp, ramus, limited bone stock Part # 487.942 with lot numbers: 3769706, 3779571, 3788741, 7806096, 7823464. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however there

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Synthes USA HQ, Inc.
Distributor
US Nationwide Distribution including the states of MA, TX, MN, SC, UT, VA, and NY.
Lot
Titanium Multi Vector Distractor Pin Holding Clamp, body, limited bone stock Part # 487.941 with lot numbers: 3769720, 3779569, 3788731, 3794434, 7806093, 7823463 and Titanium Multi Vector Distractor Pin Holding Clamp, ramus, limited bone stock Part # 487.942 with lot numbers: 3769706, 3779571, 3788741, 7806096, 7823464.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone lengthening, including conditions such as congenital mandibular deficiencies or post-traumatic defects. The Multi Vector Distractor can be utilized to perform bone transport procedures as an alternative to bone grafts and free flaps. recalled?

Yes. FDA published recall Z-1745-2013 on July 24, 2013. Reported hazard: Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however there

What should I do if I have this product?

Voluntary: Firm initiated · Visit