Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System & Brilliance CT Big Bore Radiology Whole Body Computed Tomography X-Ray System. Product Usage: The Brilliance CT Big Bore is a Whole Body Computed Tomography X-Ray System which produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.

Published July 2, 2014 · Recall initiated: December 19, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During internal testing, the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48MHZ frequency, the testing indicated the Brilliance CT Big Bore was 3.5dB uV/meter higher than the applicable IEC 60601-1-2 standard specification.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - US (nationwide) in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WV. and countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, France, French Polynesia, Gabon, Germany, India, Indonesia, Israel, Italy, Jordan, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Oman, Palestine, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Ukraine & United Kingdom.
Lot
Brilliance CT Big Bore Oncology: Model #728243; Serial #s: 7185, 7187-7189, 7191-7199, 7201,-7243, 7245-7251, 7253-7267, 7269, 7270, 7272-7297, 7299, 7300, 7302, 7304-7317, 7319-7374, 7376-7379, 7381, 7382, 7385, 7387-7389, 7391-7393, 7395, 7397-7401, 7403-7406, 7409-7429, 7431-7433, 7435-7438, 7440-7451, 7454-7457, 7459, 7462-7485, 7487-7509, 7511-7513, 7515-7519, 7521-7524, 7526-7528, 7530-7532, 7534-7537, 7539-7542, 7544, 7545, 7547, 7549-7556, 7558-7562, 7564-7566, 7568, 7569, 7573, 7574, 7577-7582, 7584, 7585, 7588-7593, 7596-7599, 7601-7603, 7605, 7608, 7609, 7611-7613, 7615-7618, 7620, 7622-7624, 7626-7630, 7634, 7635, 7637-7648 & 75000. Brilliance CT Big Bore Radiology: Model #728244; Serial #s: 7190, 7200, 7244, 7252, 7268, 7271, 7301, 7318, 7375, 7380, 7383, 7384, 7386, 7390, 7396, 7402, 7407, 7408, 7427, 7430, 7439, 7452, 7453, 7458, 7486, 7510, 7514, 7520, 7525, 7529, 7533, 7538, 7543, 7546, 7557, 7570-7572, 7575, 7576, 7583, 7586, 7587, 7595, 7600, 7604, 7607, 7610, 7614, 7619, 7621, 7625, 7631-7633 & 7636.
Risk classification
Class II
Area (FDA)
device
Country
United States