ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.
Published July 24, 2013 · Recall initiated: June 4, 2013
Summary
FDA records medical device Class II recall Z-1752-2013, published July 24, 2013 and initiated June 4, 2013. Recalling firm: Baxa Corporation; Status: Terminated; Reported quantity: 753 units; Lot: ExactaMix 1200 serial numbers 45004-45481 (sequentially), and ExactaMix 2400 serial numbers 40005-40335. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxa Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada, Latin America, Europe/Middle East/Africa, and Taiwan.
- Lot
- ExactaMix 1200 serial numbers 45004-45481 (sequentially), and ExactaMix 2400 serial numbers 40005-40335.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABaxa Vial Adapter, 14mm - Luer Lock, Order No. REF: 90, Quantity: 50 each, Single Use. Product Code H93890.September 18, 2013
- FDAPharmalucence Kit for the Preparation of Technetium Tc 99m Sestamibi Injection for Mycardial and BreastJuly 31, 2013
- FDALifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTSJuly 24, 2013
- FDAHydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used inJuly 24, 2013
- FDASiemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-rayJuly 24, 2013
- FDARepeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution inOctober 31, 2012
Questions about this recall
Was ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution. recalled?
Yes. FDA published recall Z-1752-2013 on July 24, 2013. Reported hazard: Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.
What should I do if I have this product?
Voluntary: Firm initiated · Letter