Back to search
FDAFDA · Food and drugs·Recall number Z-1756-2024

The Oncology Essentials Package of the Spectral CT imaging system.

Published May 8, 2024 · Recall initiated: March 29, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
Lot
Model No. 728333; UDI: (01)00884838101111(21); Serial No. 10069, 10084, 10169, 10137, 10163, 10196, 10181, 10199, 10210, 10206, 10202, 10139.
Risk classification
Class II
Area (FDA)
device
Country
United States