The Oncology Essentials Package of the Spectral CT imaging system.
Published May 8, 2024 · Recall initiated: March 29, 2024
Summary
FDA records medical device Class II recall Z-1756-2024, published May 8, 2024 and initiated March 29, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 12 units; Lot: Model No. 728333; UDI: (01)00884838101111(21); Serial No. 10069, 10084, 10169, 10137, 10163, 10196, 10181, 10199, 10210, 10206, 10202, 10139. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
- Lot
- Model No. 728333; UDI: (01)00884838101111(21); Serial No. 10069, 10084, 10169, 10137, 10163, 10196, 10181, 10199, 10210, 10206, 10202, 10139.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDADuloxetine Delayed-Release Capsules, USP, 30mg, Rx Only, (a) 90-count bottles (NDC 51991-747-90), (b)May 15, 2024
- FDAJohnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21LMay 8, 2024
Questions about this recall
Was The Oncology Essentials Package of the Spectral CT imaging system. recalled?
Yes. FDA published recall Z-1756-2024 on May 8, 2024. Reported hazard: Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter