Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Published May 8, 2024 · Recall initiated: March 14, 2024
Summary
FDA records medical device Class II recall Z-1755-2024, published May 8, 2024 and initiated March 14, 2024. Recalling firm: Johnson & Johnson Surgical Vision, Inc.; Status: Ongoing; Reported quantity: 3,428 units; Lot: Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Johnson & Johnson Surgical Vision, Inc.
- Distributor
- U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.
- Lot
- Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO,September 9, 2026
- FDADuloxetine Delayed-Release Capsules, USP, 30mg, Rx Only, (a) 90-count bottles (NDC 51991-747-90), (b)May 15, 2024
- FDABattery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1, model numbers: a. SM-20HF-Batt; b.May 8, 2024
- FDAThe Oncology Essentials Package of the Spectral CT imaging system.May 8, 2024
- FDAINNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cupMay 8, 2024
- FDATECNIS Toric II OptiBlue IOL Models ZCWJanuary 24, 2024
Questions about this recall
Was Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery. recalled?
Yes. FDA published recall Z-1755-2024 on May 8, 2024. Reported hazard: Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
What should I do if I have this product?
Voluntary: Firm initiated · Letter