Back to search
FDAFDA · Food and drugs·Recall number Z-1764-2024

DURAFLOW 2, 24CM STRAIGHT VASC-PAK PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012185

Published May 15, 2024 · Recall initiated: March 18, 2024

Summary

FDA records medical device Class II recall Z-1764-2024, published May 15, 2024 and initiated March 18, 2024. Recalling firm: Angiodynamics, Inc.; Status: Ongoing; Reported quantity: 83 units; Lot: UDI-DI: 25051684014820 Lot Numbers: 5805849. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
VA
Lot
UDI-DI: 25051684014820 Lot Numbers: 5805849
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DURAFLOW 2, 24CM STRAIGHT VASC-PAK PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103012185 recalled?

Yes. FDA published recall Z-1764-2024 on May 15, 2024. Reported hazard: Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.

What should I do if I have this product?

Voluntary: Firm initiated · Letter