VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK
Published April 19, 2017 · Recall initiated: March 7, 2017
Summary
FDA records medical device Class II recall Z-1765-2017, published April 19, 2017 and initiated March 7, 2017. Recalling firm: ORTHO-CLINICAL DIAGNOSTICS; Status: Terminated; Reported quantity: US: 6324; Foreign: 1868; Lot: Lot 1570 (expiry 22 March 2017), Lot 1580 (expiry 13 April 2017), Lot 1590 (expiry 27 April 2017). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Distributor
- Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.
- Lot
- Lot 1570 (expiry 22 March 2017), Lot 1580 (expiry 13 April 2017), Lot 1590 (expiry 27 April 2017)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDACobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xxApril 19, 2017
- FDAThe PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gainingApril 19, 2017
- FDAiConnect Enterprise Archive (ICEA) software. The firm name on the labeling is Merge Healthcare, Hartland, WI.April 19, 2017
- FDAVITROS Immunodiagnostic Products TSH Reagent Pack, REF/Catalog Number/Product Code 191 2997, Unique DeviceAugust 24, 2016
- FDAVITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin GOctober 1, 2014
- FDAVITROS® Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS® CA 19-9 ReagentAugust 20, 2014
Questions about this recall
Was VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK recalled?
Yes. FDA published recall Z-1765-2017 on April 19, 2017. Reported hazard: Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit