FDAFDA · Food and drugs·Recall number Z-1779-2019

Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity Sterilizer incorporates high-pressure steam to sterilize nonporous or porous items and materials used in the health care industry

Published June 19, 2019 · Recall initiated: April 18, 2019

Summary

FDA records medical device Class II recall Z-1779-2019, published June 19, 2019 and initiated April 18, 2019. Recalling firm: Skytron, Div. The KMW Group, Inc; Status: Terminated; Reported quantity: 5 units; Lot: Serial Numbers: 14092619, 14092620, 14102621, 17123520, 18113521. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Skytron, Div. The KMW Group, Inc
Distributor
US Nationwide Distribution - AK, IA, MS
Lot
Serial Numbers: 14092619, 14092620, 14102621, 17123520, 18113521
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity Sterilizer incorporates high-pressure steam to sterilize nonporous or porous items and materials used in the health care industry recalled?

Yes. FDA published recall Z-1779-2019 on June 19, 2019. Reported hazard: Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure

What should I do if I have this product?

Voluntary: Firm initiated · Letter