EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table
Published June 8, 2016 · Recall initiated: April 6, 2016
Summary
FDA records medical device Class II recall Z-1844-2016, published June 8, 2016 and initiated April 6, 2016. Recalling firm: Skytron, Div. The KMW Group, Inc; Status: Terminated; Reported quantity: 458; Lot: Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Skytron, Div. The KMW Group, Inc
- Distributor
- Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.
- Lot
- Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the SkyVision Linx 300October 14, 2020
- FDASkytron Ergon 3 Series Skyboom Monitor Bracket-Flatscreen Bracket, single monitor mount, 27" max. screen sizeApril 1, 2020
- FDASkytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON IntegrityJune 19, 2019
- FDACOSEAL Surgical Sealant Kit, 8 mL, Product Code: 934072; For use in vascular reconstructions to achieveJune 8, 2016
- FDACurity All Purpose Sponges, Non-woven, 4 Ply, Soft Pouch, 4" x 4" (10.2 cm x 10.2 cm) Item Number: 8047June 8, 2016
- FDACOSEAL Surgical Sealant Kit, 4 mL, Product Code: 934071; For use in vascular reconstructions to achieveJune 8, 2016
Questions about this recall
Was EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table recalled?
Yes. FDA published recall Z-1844-2016 on June 8, 2016. Reported hazard: The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
What should I do if I have this product?
Voluntary: Firm initiated · Letter