Rultract Retractor Ratchet Cover Plate Distributed by Pemco, Inc., Cleveland, OH 44131 Surgical retractor used to lift the incision site

Published July 31, 2013 · Recall initiated: July 10, 2013

Summary

FDA records medical device Class II recall Z-1783-2013, published July 31, 2013 and initiated July 10, 2013. Recalling firm: Pemco Inc; Status: Terminated; Reported quantity: 87 units; Lot: M/N 4100-IMR-6: S/N's: 4829-4839, 4845-4851, 4857-4871, 4874-4884, 4894-4908 & 4916-4923; M/N 4100-IMR-41: S/N's: 4840-4844, 4852-4856, 4888-4893, 4913-4915 & 4926-4927. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm became aware of the problem when a ratchet cover plate on a device was returned for repair showed signs of corrosion.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Pemco Inc
Distributor
Worldwide Distribution - USA including AL, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NJ, NM, NV, OK, PA, TN, TX, UT, VA & WA. Internationally to Czech Republic, Italy, Japan, Korea & Singapore.
Lot
M/N 4100-IMR-6: S/N's: 4829-4839, 4845-4851, 4857-4871, 4874-4884, 4894-4908 & 4916-4923; M/N 4100-IMR-41: S/N's: 4840-4844, 4852-4856, 4888-4893, 4913-4915 & 4926-4927.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Rultract Retractor Ratchet Cover Plate Distributed by Pemco, Inc., Cleveland, OH 44131 Surgical retractor used to lift the incision site recalled?

Yes. FDA published recall Z-1783-2013 on July 31, 2013. Reported hazard: The firm became aware of the problem when a ratchet cover plate on a device was returned for repair showed signs of corrosion.

What should I do if I have this product?

Voluntary: Firm initiated · Letter