GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.
Published June 20, 2012 · Recall initiated: May 31, 2006
Summary
FDA records medical device Class II recall Z-1789-2012, published June 20, 2012 and initiated May 31, 2006. Recalling firm: General Electric Med Systems LLC; Status: Terminated; Reported quantity: 3,796 units; Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- General Electric Med Systems LLC
- Distributor
- Worldwide Distribution.
- Lot
- All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGE Healthcare Revolution XR/d; GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model numbers:June 20, 2012
- FDAGE Proteus XR/a Radiographic System. GE Medical Systems-Americas Milwaukee, WI. Model Number: 2259988. ProductJune 20, 2012
- FDAPhilips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- NationwideJune 20, 2012
- FDAThe Brilliance iCT and iCT SP are Computed Tomography X-Ray Systems intended to produce cross-sectional imagesJune 20, 2012
- FDASigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 --June 20, 2012
- FDALipofilter 3000, model ASP-CAN-2, labeled in part ***MicroAire, 1641 Edlich Drive, Charlottesville, VA 22911,June 20, 2012
Questions about this recall
Was GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use. recalled?
Yes. FDA published recall Z-1789-2012 on June 20, 2012. Reported hazard: An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
What should I do if I have this product?
Voluntary: Firm initiated · Letter