RKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component
Published May 22, 2024 · Recall initiated: March 25, 2024
Summary
FDA records medical device Class II recall Z-1790-2024, published May 22, 2024 and initiated March 25, 2024. Recalling firm: Ossur H / F; Status: Ongoing; Reported quantity: 1 unit; Lot: UDI: 05690977467811/ Serial Number: HF363264. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ossur H / F
- Distributor
- US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
- Lot
- UDI: 05690977467811/ Serial Number: HF363264
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Iceland
Related recalls
- FDAMiami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101October 9, 2024
- FDARKNXC0003, RHEO KNEE XC 3 YR WARRANTY, MD, CE, External lower limb prosthetic componentMay 22, 2024
- FDARKN130003, RHEO KNEE 3 YR WARRANTY, MD, CE, External lower limb prosthetic componentMay 22, 2024
- FDARKN130002, RHEO KNEE 2 YR WARRANTY, MD, CE, External lower limb prosthetic componentMay 22, 2024
- FDATriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM NailMay 22, 2024
- FDATriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM NailMay 22, 2024
Questions about this recall
Was RKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component recalled?
Yes. FDA published recall Z-1790-2024 on May 22, 2024. Reported hazard: Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.
What should I do if I have this product?
Voluntary: Firm initiated · Letter