Back to search
FDAFDA · Food and drugs·Recall number Z-1790-2024

RKNXC0005, RHEO KNEE XC 5 YR WARRANTY, MD, CE, External lower limb prosthetic component

Published May 22, 2024 · Recall initiated: March 25, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ossur H / F
Distributor
US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Lot
UDI: 05690977467811/ Serial Number: HF363264
Risk classification
Class II
Area (FDA)
device
Country
Iceland