TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only
Published May 22, 2024 · Recall initiated: March 19, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- TriMed Inc.
- Distributor
- US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
- Lot
- UDI:00842188127885/ Lot # QM23087
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States