Discovery NM/CT 670 dual detector free-geometry integrated nuclear imaging system. Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.
Published July 31, 2013 · Recall initiated: June 13, 2013
Summary
FDA records medical device Class I recall Z-1792-2013, published July 31, 2013 and initiated June 13, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 213; Lot: System ID: 0828260021, 0828260022, 0850260118, 0850260122, 0850260128, 0850260131, 0850260132, 0853260094, 0853260099, 0856260033, 0856260034, 0856260035, 0910262297, 0910262304, 0910262310, 0910262319, 0910262322, 0910262325, 0910262344, 0910263223, 0910263228, 0910263231, 0910263235, 0910263239, 0910263241, 0910263243, 0910263250, 0910264135, 0910264141, 0910264147, 0910265074, 0910265075, 0910265076, 0920269097, 082426040043, 082426040044, 082426060005, 082426070028, 082426070030, 082426080020, 082426090029, 082426090030, 082426130029, 082426130030, 082426150026, 082426160042, 082426210020, 082426230020, 083026009408410, 083026023683711, 083026201006810, 00001NUC12, 00028NUC01, 00042NUC02, 00046NUC06, 00065NUC08, 00123NUC01, 00149NUC01, 00149NUC03, 00197NUC14, 00199NUC13, 00200NUC07, 00200NUC08, 00234NUC09, 00361NUC08, 00432NUC02, 010120NU19, 030037NU08, 030082NU13, 030133NU13, 030360NU01, 05462NUC05, 06007NUC19, 080041NU04, 100032NU05, 10497NUC20, 110017NU06, 110022NU05, 11018NUC08, 190052NU09, 190163NU04, 190163NU06, 204726D670NM, 210060NU02, 212639NMD670, 250370D670, 260010NU06, 262574D670, 305674D670, 309671SPECT, 403343D670, 403943D670, 403943D670B, 404686NMCT, 410955D670, 410D6701, 415353D670, 416495D670, 418549D670A, 418549D670B, 423495D670, 514281D670, 514338D670, 514842AD670, 514848D670, 514876D670, 519973D670, 529150NU05, 540332D670, 600092NM11, 610402D670NM1, 613548D670, 614NCH670, 615322NM5, 615893MMCT1, 617638DSV670, 630432D670, 720848D670, 803751D670, 819563D670, 819563D670B, 856247D670, 856247NM, 900010NU09, 905472D670, 905684D670A, 905684D670B, 905883D670, 972747NM670, 972935D670, A4169312, A4194844, A51051127, A51142126, A5125139, A5129983, A5196344, B4182011, B5140618, BG4009NM01, D007NU12, DK1029NU14, DK1047NM01, EE1020NM03, FE416061, FI1052MN01, GP1000NM01, HOR0281, IL1020NM36, IL1041NM30, IL1062NM29, M001NU42, M2002713, M251400603, M4134637, M4143630, M4143993, M4144126, M4145158, M4194042, M4198090, M4210246, M4215577, M4476424, M4505290, M5172553, M5400562, M5480777, M9409545, M9886847, MEX68915, NO1018MN06, NO1087MN09, NO1088MN16, NT1001, NT1002, NT1003, NT1004, NT1005, NT1006, NT1007, NT1008, NT1009, NT1010, NT1011, NT1013, NT1015, NT1020, NT2001, NT2002, O005NU27, PC8262NU01, PITT670, RTD0737, RU1095NM03, RU1279NM02, RU1949NM01, RU3451NM01, RU3467NM01, RU3623NM01, RU3637PT01, RU3713NM01, RU3900NM01, RU6340NM01, RU7175NM01, RU7187NM01, RU7505NM01, UX2305NU01, V4144610, V4169306, X102874005, X4144501, XM203402. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Venezuela, Vietnam.
- Lot
- System ID: 0828260021, 0828260022, 0850260118, 0850260122, 0850260128, 0850260131, 0850260132, 0853260094, 0853260099, 0856260033, 0856260034, 0856260035, 0910262297, 0910262304, 0910262310, 0910262319, 0910262322, 0910262325, 0910262344, 0910263223, 0910263228, 0910263231, 0910263235, 0910263239, 0910263241, 0910263243, 0910263250, 0910264135, 0910264141, 0910264147, 0910265074, 0910265075, 0910265076, 0920269097, 082426040043, 082426040044, 082426060005, 082426070028, 082426070030, 082426080020, 082426090029, 082426090030, 082426130029, 082426130030, 082426150026, 082426160042, 082426210020, 082426230020, 083026009408410, 083026023683711, 083026201006810, 00001NUC12, 00028NUC01, 00042NUC02, 00046NUC06, 00065NUC08, 00123NUC01, 00149NUC01, 00149NUC03, 00197NUC14, 00199NUC13, 00200NUC07, 00200NUC08, 00234NUC09, 00361NUC08, 00432NUC02, 010120NU19, 030037NU08, 030082NU13, 030133NU13, 030360NU01, 05462NUC05, 06007NUC19, 080041NU04, 100032NU05, 10497NUC20, 110017NU06, 110022NU05, 11018NUC08, 190052NU09, 190163NU04, 190163NU06, 204726D670NM, 210060NU02, 212639NMD670, 250370D670, 260010NU06, 262574D670, 305674D670, 309671SPECT, 403343D670, 403943D670, 403943D670B, 404686NMCT, 410955D670, 410D6701, 415353D670, 416495D670, 418549D670A, 418549D670B, 423495D670, 514281D670, 514338D670, 514842AD670, 514848D670, 514876D670, 519973D670, 529150NU05, 540332D670, 600092NM11, 610402D670NM1, 613548D670, 614NCH670, 615322NM5, 615893MMCT1, 617638DSV670, 630432D670, 720848D670, 803751D670, 819563D670, 819563D670B, 856247D670, 856247NM, 900010NU09, 905472D670, 905684D670A, 905684D670B, 905883D670, 972747NM670, 972935D670, A4169312, A4194844, A51051127, A51142126, A5125139, A5129983, A5196344, B4182011, B5140618, BG4009NM01, D007NU12, DK1029NU14, DK1047NM01, EE1020NM03, FE416061, FI1052MN01, GP1000NM01, HOR0281, IL1020NM36, IL1041NM30, IL1062NM29, M001NU42, M2002713, M251400603, M4134637, M4143630, M4143993, M4144126, M4145158, M4194042, M4198090, M4210246, M4215577, M4476424, M4505290, M5172553, M5400562, M5480777, M9409545, M9886847, MEX68915, NO1018MN06, NO1087MN09, NO1088MN16, NT1001, NT1002, NT1003, NT1004, NT1005, NT1006, NT1007, NT1008, NT1009, NT1010, NT1011, NT1013, NT1015, NT1020, NT2001, NT2002, O005NU27, PC8262NU01, PITT670, RTD0737, RU1095NM03, RU1279NM02, RU1949NM01, RU3451NM01, RU3467NM01, RU3623NM01, RU3637PT01, RU3713NM01, RU3900NM01, RU6340NM01, RU7175NM01, RU7187NM01, RU7505NM01, UX2305NU01, V4144610, V4169306, X102874005, X4144501, XM203402.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDADiscovery NM 630, dual detector free-geometry integrated nuclear imaging system. Product Usage: The GEJuly 31, 2013
- FDAOptima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear MedicineJuly 31, 2013
Questions about this recall
Was Discovery NM/CT 670 dual detector free-geometry integrated nuclear imaging system. Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. recalled?
Yes. FDA published recall Z-1792-2013 on July 31, 2013. Reported hazard: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
What should I do if I have this product?
Voluntary: Firm initiated · Letter