PlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System) The PlasmaJet system is a neutral plasma surgery system that is designed for cutting and coagulation in open surgery and laparoscopic surgery.
Published June 20, 2012 · Recall initiated: April 27, 2012
Summary
FDA records medical device Class II recall Z-1794-2012, published June 20, 2012 and initiated April 27, 2012. Recalling firm: Plasma Surgical Inc.; Status: Terminated; Reported quantity: 36 units; Lot: Lot Number 40041, exp. date February 2015 and Lot Number 40043, exp. date February 2015. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential failure in the glued seal between the stem and the tip of the handpiece leading to the possible leakage of a very small drop of distilled water coolant from the handpiece in use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Plasma Surgical Inc.
- Distributor
- Worldwide distribution: USA (nationwide) including states of: CA and NY; and country of: United Kingdom.
- Lot
- Lot Number 40041, exp. date February 2015 and Lot Number 40043, exp. date February 2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAColumbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus RevisionJune 20, 2012
- FDAIntegra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1)June 20, 2012
- FDALipofilter 3000, model ASP-CAN-2, labeled in part ***MicroAire, 1641 Edlich Drive, Charlottesville, VA 22911,June 20, 2012
- FDAIntegra Mozaik Putty 2.5cc, 5 cc, 10 cc Product Usage: The product is a bone graft substitute composed ofJune 20, 2012
- FDAThe Brilliance iCT and iCT SP are Computed Tomography X-Ray Systems intended to produce cross-sectional imagesJune 20, 2012
- FDAVygon VY(TM)Flush Heparin Lock Flush Syringes, 1 Unit per mL, 3mL in a 10mL Flush Syringe, 30 syringes per boxJune 20, 2012
Questions about this recall
Was PlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System) The PlasmaJet system is a neutral plasma surgery system that is designed for cutting and coagulation in open surgery and laparoscopic surgery. recalled?
Yes. FDA published recall Z-1794-2012 on June 20, 2012. Reported hazard: Potential failure in the glued seal between the stem and the tip of the handpiece leading to the possible leakage of a very small drop of distilled water coolant from the handpiece in use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter