Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.
Published July 31, 2013 · Recall initiated: June 5, 2013
Summary
FDA records medical device Class II recall Z-1795-2013, published July 31, 2013 and initiated June 5, 2013. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 401; Lot: Model Number 5895003 with serial numbers 1052 through 2500. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Nationwide Distribution.
- Lot
- Model Number 5895003 with serial numbers 1052 through 2500.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDATotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax QuantumJuly 31, 2013
- FDATEG 5000 Thromboelastograph Coagulation Analyzer Product Usage: Multipurpose System for In Vitro CoagulationJuly 31, 2013
Questions about this recall
Was Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging. recalled?
Yes. FDA published recall Z-1795-2013 on July 31, 2013. Reported hazard: Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
What should I do if I have this product?
Voluntary: Firm initiated · Letter