TotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax Quantum Convertible mattress The TotalCare Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Published July 31, 2013 · Recall initiated: May 30, 2013
Summary
FDA records medical device Class II recall Z-1794-2013, published July 31, 2013 and initiated May 30, 2013. Recalling firm: Hill-Rom, Inc.; Status: Terminated; Reported quantity: 4,121 Units; Lot: Bed model numbers: PAU001010784, PAU005010784, P1900, PAK169N10184, PAK169N10384, PA8171210184, PA8175210184, PA8175210384, PA8171210384, PA8171310384, and PA8173210384. The AccuMax Quantum VPC, Complete and Convertible mattresses are not serialized, but were distributed between September 2005 and April 2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the bed is put into the chair egress position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the bed at an angle. A patient attempting to get back into the bed may miss the seat and sit on the foot section instead. The patient can slide down to the floor, or fall, with possible injury.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hill-Rom, Inc.
- Distributor
- Worldwide distribution: USA (nationwide) and countries of: Canada, Colombia, Nigeria, Kuwait and Saudi Arabia.
- Lot
- Bed model numbers: PAU001010784, PAU005010784, P1900, PAK169N10184, PAK169N10384, PA8171210184, PA8175210184, PA8175210384, PA8171210384, PA8171310384, and PA8173210384. The AccuMax Quantum VPC, Complete and Convertible mattresses are not serialized, but were distributed between September 2005 and April 2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProgressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or otherAugust 5, 2020
- FDASteriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.July 1, 2015
- FDAViking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.June 24, 2015
- FDAAffinity® Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of childSeptember 10, 2014
- FDAOptima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear MedicineJuly 31, 2013
- FDAAxiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used forJuly 31, 2013
Questions about this recall
Was TotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax Quantum Convertible mattress The TotalCare Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution. recalled?
Yes. FDA published recall Z-1794-2013 on July 31, 2013. Reported hazard: When the bed is put into the chair egress position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the bed at an angle. A patient attempting to get back into the bed may miss the seat and sit on the foot section instead. The patient can slide down to the floor, or fall, with possible injury.
What should I do if I have this product?
Voluntary: Firm initiated · Letter