VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.
Published June 25, 2014 · Recall initiated: May 12, 2014
Summary
FDA records medical device Class II recall Z-1804-2014, published June 25, 2014 and initiated May 12, 2014. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: Domestic - 612 units, Foreign - 461 units; Lot: Lots 1732 (exp. 31-Jul-2014), 1743 (exp. 31-Jul-2014), 1753 (exp. 28-Feb-2015), 1764 (31-Jul-2015). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
- Lot
- Lots 1732 (exp. 31-Jul-2014), 1743 (exp. 31-Jul-2014), 1753 (exp. 28-Feb-2015), 1764 (31-Jul-2015)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAEZPass Passer, Sterile single use arthroscopic instrument utilized in conjunction with a nitinol wire to aidJune 25, 2014
- FDAVITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnosticJune 25, 2014
Questions about this recall
Was VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent. recalled?
Yes. FDA published recall Z-1804-2014 on June 25, 2014. Reported hazard: Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.
What should I do if I have this product?
Voluntary: Firm initiated · Letter