Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system
Published June 27, 2012 · Recall initiated: May 22, 2012
Summary
FDA records medical device Class II recall Z-1818-2012, published June 27, 2012 and initiated May 22, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 752 units; Lot: Software release 2.0.2 and 2.0.2SP1 including mirror icon. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Canada AFGHANISTAN ALBANIA ALGERIA ANGOLA ARGENTINA ARMENIA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BAHRAIN BANGLADESH BARBADOS BELARUS BELGIUM BENIN BHUTAN BOLIVIA BOSNIA&HERZEGOVINA BRAZIL BRUNEI DARUSSALAM BULGARIA BURUNDI CAMBODIA CAMEROON CANARY ISLANDS CHAD CHILE CHINA COLOMBIA CONGO CONGO,DEMOCRATIC COSTA RICA COTE D'IVOIRE CROATIA CUBA CYPRUS CZECH REPUBLIC DENMARK DJIBOUTI DOMINICAN REPUBLIC ECUADOR EGYPT EL SALVADOR EQUATORIAL GUINEA ERITREA ESTONIA ETHIOPIA fake test ssd FINLAND FRANCE FRENCH GUIANA GABON GAMBIA GEORGIE GERMANY GHANA GREECE GUADELOUPE GUAM GUATEMALA GUINEA GUYANA HONDURAS HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAMAICA JAPAN JORDAN KAZAKSTAN KENYA KUWAIT KYRGYZSTAN LAO PEOPLE'S DEM.REP LATVIA LEBANON LIBERIA LIBYAN ARAB JAMAHIRI LIECHTENSTEIN LITHUANIA LUXEMBOURG MACAU MACEDONIA MALAWI MALAYSIA MALI MALTA MARTINIQUE MAURITANIA MAURITIUS METRACOM MEXICO MOLDOVA MONGOLIA MOROCCO MOZAMBIQUE NAMIBIA NEPAL NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NIGERIA NORTH KOREA NORWAY OMAN PAKISTAN PALESTINE PANAMA PARAGUAY PERU PHC BEST PHILIPPINES PMCC BU Canada PMCC BU USA POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA&MONTE NEGRO SIERRA LEONE SINGAPORE SLOVAKIA SLOVENIA SOMALIA SOUTH AFRICA SOUTH KOREA SPAIN SRI LANKA SUDAN SURINAME SWEDEN SWITZERLAND SYRIAN ARAB REPUBLIC TAIWAN TAJIKISTAN TANZANIA THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY TURKMENISTAN UGANDA UKRAINE Unidentified Country UNITED ARAB EMIRATES UNITED KINGDOM URUGUAY UZBEKISTAN VENEZUELA VIET NAM YEMEN ZAMBIA ZIMBABWE
- Lot
- Software release 2.0.2 and 2.0.2SP1 including mirror icon
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDATX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator thatJune 27, 2012
- FDAUS-001 Rev A brochure is a two sided tri-fold printed marketing brochure describing the approved labelingJune 27, 2012
Questions about this recall
Was Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system recalled?
Yes. FDA published recall Z-1818-2012 on June 27, 2012. Reported hazard: With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side
What should I do if I have this product?
Voluntary: Firm initiated · Letter