Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039
Published April 26, 2017 · Recall initiated: March 22, 2017
Summary
FDA records medical device Class II recall Z-1820-2017, published April 26, 2017 and initiated March 22, 2017. Recalling firm: Philips Electronics North America Corporation; Status: Terminated; Reported quantity: 6992 (total for all devices); Lot: Devices with Laird Chiller. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Electronics North America Corporation
- Distributor
- Nationwide, worldwide including Canada
- Lot
- Devices with Laird Chiller.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAllura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications,October 3, 2018
- FDANuGo Slim Crunchy Peanut Butter Protein Bar, 5g net carbs, A gluten-free, soy protein bar with overall peanutApril 26, 2017
- FDAThe following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura XperApril 26, 2017
Questions about this recall
Was Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039 recalled?
Yes. FDA published recall Z-1820-2017 on April 26, 2017. Reported hazard: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What should I do if I have this product?
Voluntary: Firm initiated · Letter