FDAFDA · Food and drugs·Recall number Z-1820-2017
Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039
Published April 26, 2017 · Recall initiated: March 22, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Electronics North America Corporation
- Distributor
- Nationwide, worldwide including Canada
- Lot
- Devices with Laird Chiller.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States