The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031
Published April 26, 2017 · Recall initiated: March 22, 2017
Summary
FDA records medical device Class II recall Z-1822-2017, published April 26, 2017 and initiated March 22, 2017. Recalling firm: Philips Electronics North America Corporation; Status: Terminated; Reported quantity: 6992 (total distributed for all products); Lot: Devices with a Laird Chiller. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Electronics North America Corporation
- Distributor
- Nationwide, worldwide including Canada
- Lot
- Devices with a Laird Chiller.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAllura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications,October 3, 2018
- FDAAllura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10April 26, 2017
- FDATapered SwissPlus Implant System, Model No. SPB10April 26, 2017
Questions about this recall
Was The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031 recalled?
Yes. FDA published recall Z-1822-2017 on April 26, 2017. Reported hazard: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What should I do if I have this product?
Voluntary: Firm initiated · Letter