Back to search
FDAFDA · Food and drugs·Recall number Z-1831-2014

Coulter DxH Cleaner 10L, Part No. 628023. For use on UniCel DxH Systems for components that come in contact with blood samples.

Published July 2, 2014 · Recall initiated: November 8, 2013

Summary

FDA records medical device Class II recall Z-1831-2014, published July 2, 2014 and initiated November 8, 2013. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 12,924 units total (9,733 units in US); Lot: Lot No. 3040210 3040330 3043090 3043100 3057970 3057980 3058130 3071440 3071450. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution -- US, Canada, Australia, China, Belgium, Brunei Darssalam, Colombia, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Macao, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Panama, Portugal, Puerto Rico, Qatar, Romania, Russia Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Lot
Lot No. 3040210 3040330 3043090 3043100 3057970 3057980 3058130 3071440 3071450
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Coulter DxH Cleaner 10L, Part No. 628023. For use on UniCel DxH Systems for components that come in contact with blood samples. recalled?

Yes. FDA published recall Z-1831-2014 on July 2, 2014. Reported hazard: Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte.

What should I do if I have this product?

Voluntary: Firm initiated · Letter