The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table.
Published July 9, 2014 · Recall initiated: March 15, 2014
Summary
FDA records medical device Class II recall Z-1835-2014, published July 9, 2014 and initiated March 15, 2014. Recalling firm: Claymount Assembli; Status: Terminated; Reported quantity: 125 units shipped to United States; Lot: *** US Devices Only *** 1) Part ID: 17007; Serial Numbers: WO171240/E-1, WO171240/E-2. 2) Part ID: 17009; Serial Numbers: WO165307/E-1 , WO165308/E-1 , WO170263/E-1 , WO172341/E-1 , WO172341/E-1 , WO172760/E-1 , WO172760/E-2 , WO174533/E-2 , WO174980/E-2 , WO175009/E-1 , WO175010/E-2 , WO175011/E-1 , WO175011/E-2 , WO175013/E-1 , WO175013/E-2 , WO177682/E-1 , WO187248/E-1 , WO191933/E-2. 3) Part ID: 17675; Serial Numbers: WO160580/E-1 , WO166945/E-1 , WO166945/E-2 , WO166946/E-1 , WO166946/E-2 , WO166949/E-1 , WO166949/E-2 , WO166950/E-1 , WO166951/E-1 , WO166952/E-1 , WO166952/E-2 , WO166954/E-1 , WO166954/E-2 , WO166955/E-1 , WO166955/E-2 , WO166957/E-1 , WO166957/E-2 , WO169388/E-1 , WO169388/E-2 , WO172050/E-1 , WO172050/E-2 , WO172051/E-1 , WO172051/E-2 , WO172345/E-1 , WO174072/E-2 , WO175014/E-1 , WO175014/E-2 , WO175015/E-1 , WO175993/E-2 , WO179503/E-1 , WO182220/E-1. 4) Part ID: 17676; Serial Numbers: WO160452/E-1 , WO160453/E-1 , WO165325/E-1 , WO166958/E-1 , WO166958/E-2 , WO166961/E-1 , WO166962/E-1 , WO166962/E-2 , WO166963/E-1 , WO166963/E-2 , WO166964/E-1 , WO166968/E-1 , WO166968/E-1 , WO166968/E-2 , WO166969/E-1 , WO166970/E-1 , WO166970/E-2 , WO167440/E-1 , WO167440/E-2 , WO169398/E 2 , WO169398/E-1 , WO169400/E-1 , WO169400/E-2 , WO169402/E-1 , WO171077/E-1 , WO171077/E-2 , WO171390/E-1 , WO171390/E-2 , WO171391/E-1 , WO171391/E-2 , WO172052/E-1 , WO172052/E-2 , WO172054/E-1 , WO172054/E-2 , WO174099/E-2 , WO175166/E-1 , WO175166/E-2 , WO175169/E-1 , WO175169/E-2. 5) Part ID: 17985; Serial Numbers: WO166912/E-1 , WO166912/E-2 , WO166913/E-1 , WO166913/E-2 , WO169337/E-1 , WO169337/E-2 , WO169338/E-1 , WO172968/E-1 , WO172968/E-2 , WO172969/E-1 , WO172969/E-2 , WO175175/E-1 . 6) Part ID: 18423; Serial Numbers: WO193031/E-1 , WO193031/E-2 , WO193032/E-1 , WO193032/E-2 , WO196488/U-1 , WO196488/U-2 , WO196488/U-3 , WO198927/E-1 , WO198927/E-2 , WO200261/U-1. 7) Part ID: 18424; Serial Numbers: WO193035/E-1 , WO193035/E-2 , WO193036/E-1 , WO193040/E-1 , WO193040/E-2 , WO193041/E-1 , WO196489/U-1 , WO196489/U-2 , WO196489/U-3. 8) Part ID: 18425; Serial Numbers: WO193039/E-1 , WO200259/U-1 , WO200259/U-2 , WO200906/E-1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Harmony1 models of the Harmony Bucky is incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Claymount Assembli
- Distributor
- Worldwide Distribution - US including the states of CA, CO, FL, IL, NC, NJ, OH and PA and the countries of Canada, France, Italy, Netherlands, South Africa, Switzerland, Turkey and United Kingdom.
- Lot
- *** US Devices Only *** 1) Part ID: 17007; Serial Numbers: WO171240/E-1, WO171240/E-2. 2) Part ID: 17009; Serial Numbers: WO165307/E-1 , WO165308/E-1 , WO170263/E-1 , WO172341/E-1 , WO172341/E-1 , WO172760/E-1 , WO172760/E-2 , WO174533/E-2 , WO174980/E-2 , WO175009/E-1 , WO175010/E-2 , WO175011/E-1 , WO175011/E-2 , WO175013/E-1 , WO175013/E-2 , WO177682/E-1 , WO187248/E-1 , WO191933/E-2. 3) Part ID: 17675; Serial Numbers: WO160580/E-1 , WO166945/E-1 , WO166945/E-2 , WO166946/E-1 , WO166946/E-2 , WO166949/E-1 , WO166949/E-2 , WO166950/E-1 , WO166951/E-1 , WO166952/E-1 , WO166952/E-2 , WO166954/E-1 , WO166954/E-2 , WO166955/E-1 , WO166955/E-2 , WO166957/E-1 , WO166957/E-2 , WO169388/E-1 , WO169388/E-2 , WO172050/E-1 , WO172050/E-2 , WO172051/E-1 , WO172051/E-2 , WO172345/E-1 , WO174072/E-2 , WO175014/E-1 , WO175014/E-2 , WO175015/E-1 , WO175993/E-2 , WO179503/E-1 , WO182220/E-1. 4) Part ID: 17676; Serial Numbers: WO160452/E-1 , WO160453/E-1 , WO165325/E-1 , WO166958/E-1 , WO166958/E-2 , WO166961/E-1 , WO166962/E-1 , WO166962/E-2 , WO166963/E-1 , WO166963/E-2 , WO166964/E-1 , WO166968/E-1 , WO166968/E-1 , WO166968/E-2 , WO166969/E-1 , WO166970/E-1 , WO166970/E-2 , WO167440/E-1 , WO167440/E-2 , WO169398/E 2 , WO169398/E-1 , WO169400/E-1 , WO169400/E-2 , WO169402/E-1 , WO171077/E-1 , WO171077/E-2 , WO171390/E-1 , WO171390/E-2 , WO171391/E-1 , WO171391/E-2 , WO172052/E-1 , WO172052/E-2 , WO172054/E-1 , WO172054/E-2 , WO174099/E-2 , WO175166/E-1 , WO175166/E-2 , WO175169/E-1 , WO175169/E-2. 5) Part ID: 17985; Serial Numbers: WO166912/E-1 , WO166912/E-2 , WO166913/E-1 , WO166913/E-2 , WO169337/E-1 , WO169337/E-2 , WO169338/E-1 , WO172968/E-1 , WO172968/E-2 , WO172969/E-1 , WO172969/E-2 , WO175175/E-1 . 6) Part ID: 18423; Serial Numbers: WO193031/E-1 , WO193031/E-2 , WO193032/E-1 , WO193032/E-2 , WO196488/U-1 , WO196488/U-2 , WO196488/U-3 , WO198927/E-1 , WO198927/E-2 , WO200261/U-1. 7) Part ID: 18424; Serial Numbers: WO193035/E-1 , WO193035/E-2 , WO193036/E-1 , WO193040/E-1 , WO193040/E-2 , WO193041/E-1 , WO196489/U-1 , WO196489/U-2 , WO196489/U-3. 8) Part ID: 18425; Serial Numbers: WO193039/E-1 , WO200259/U-1 , WO200259/U-2 , WO200906/E-1.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAThe Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationaryJuly 9, 2014
- FDAYangban Rice Porridge with Abalone, Net Wt. 10.16 oz (288 g), UPC 8 83298 61617 3, 24 cans/caseJuly 9, 2014
- FDAOcean's Catch Jonah Crab Minced Meat Crabmeat , Net WT. 10 lb. plastic bagJuly 9, 2014
- FDAMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. IntendedJuly 9, 2014
- FDAOcean's Catch RTE All Natural Jonah Crab Combo Meat Fresh Hand Picked Crabmeat , Net WT. 8 oz. and 16 oz.July 9, 2014
- FDAMedtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Catalog No. 46914. IntendedJuly 9, 2014
Questions about this recall
Was The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. recalled?
Yes. FDA published recall Z-1835-2014 on July 9, 2014. Reported hazard: Harmony1 models of the Harmony Bucky is incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J.
What should I do if I have this product?
Voluntary: Firm initiated · Letter