COSEAL Surgical Sealant Kit, 8 mL, Product Code: 934072; For use in vascular reconstructions to achieve adjunctive hemostasis by mechanically sealing areas of leakage.
Published June 8, 2016 · Recall initiated: May 12, 2016
Summary
FDA records medical device Class II recall Z-1840-2016, published June 8, 2016 and initiated May 12, 2016. Recalling firm: Baxter Healthcare Corp; Status: Terminated; Reported quantity: 6,128 units; Lot: HA151038, HA151220, HA151037, HA151028, HA160114, HA160115, HA160222. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp
- Distributor
- Distributed US (nationwide) including Puerto Rico and in the Bahamas, Canada, and Singapore.
- Lot
- HA151038, HA151220, HA151037, HA151028, HA160114, HA160115, HA160222
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACOSEAL Surgical Sealant Kit, 4 mL, Product Code: 934071; For use in vascular reconstructions to achieveJune 8, 2016
- FDAEZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General PurposeJune 8, 2016
- FDAFruit Tartlets, Raspberry flavor, 9 - 0.56oz (16 g) tartlets - Net Wt. 5.1oz (145 g). Manufactured in France.June 8, 2016
- FDAHeparin Sodium and 0.9% Sodium Chloride Injection, 1000 USP Heparin Units, 500 mL in a VIAFLEX Plus Container,June 17, 2015
- FDA0.9% Sodium Chloride Injection USP, 100 mL MINI-BAG Plus Container, Rx only, Baxter Healthcare Corporation,April 15, 2015
- FDA0.9% Sodium Chloride Injection, USP, 50 mL, VIAFLEX Plastic Container Multi Pack, Rx Only, Baxter HealthcareApril 8, 2015
Questions about this recall
Was COSEAL Surgical Sealant Kit, 8 mL, Product Code: 934072; For use in vascular reconstructions to achieve adjunctive hemostasis by mechanically sealing areas of leakage. recalled?
Yes. FDA published recall Z-1840-2016 on June 8, 2016. Reported hazard: Potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter