Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866
Published October 5, 2022 · Recall initiated: August 16, 2022
Summary
FDA records medical device Class II recall Z-1840-2022, published October 5, 2022 and initiated August 16, 2022. Recalling firm: Cook Incorporated; Status: Ongoing; Reported quantity: 115 US; Lot: RLPC-35-145 G06866 UDI-DI: 00827002068667 Lots/Expiration Date: 14828631 06-07-2025 14828643 06-07-2025 14828644 06-07-2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Incorporated
- Distributor
- Domestic distribution nationwide. Foreign distribution worldwide.
- Lot
- RLPC-35-145 G06866 UDI-DI: 00827002068667 Lots/Expiration Date: 14828631 06-07-2025 14828643 06-07-2025 14828644 06-07-2025
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAWilliams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, andOctober 21, 2026
- FDACentimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographicMay 13, 2026
- FDAAurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters areMay 13, 2026
- FDABeacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20May 13, 2026
- FDANewton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701October 5, 2022
- FDARosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623October 5, 2022
Questions about this recall
Was Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866 recalled?
Yes. FDA published recall Z-1840-2022 on October 5, 2022. Reported hazard: Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
What should I do if I have this product?
Voluntary: Firm initiated · Letter