Prestige Plus 185cm J-Tip: Part number: 9185J; Product Usage: The Prestige Plus Wire is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Published July 1, 2015 · Recall initiated: June 4, 2015

Summary

FDA records medical device Class II recall Z-1845-2015, published July 1, 2015 and initiated June 4, 2015. Recalling firm: Volcano Corporation; Status: Terminated; Reported quantity: 162 total (both devices); Lot: Product number 9185J; lot number 50004551/005-060, serial numbers 005-060. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and not in accordance with the original shelf life date at the time of manufacture

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Volcano Corporation
Distributor
US Nationwide Distribution and one account in Canada.
Lot
Product number 9185J; lot number 50004551/005-060, serial numbers 005-060.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Prestige Plus 185cm J-Tip: Part number: 9185J; Product Usage: The Prestige Plus Wire is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease. recalled?

Yes. FDA published recall Z-1845-2015 on July 1, 2015. Reported hazard: During a relabeling procedure at the manufacturing facility, the products were labeled with a shelf life in accordance with the date of the relabeling and not in accordance with the original shelf life date at the time of manufacture

What should I do if I have this product?

Voluntary: Firm initiated · Letter