NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-302 Lactate , 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100, Batch Size (liters): 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch.
Published July 2, 2014 · Recall initiated: May 14, 2014
Summary
FDA records medical device Class II recall Z-1855-2014, published July 2, 2014 and initiated May 14, 2014. Recalling firm: NxStage Medical, Inc.; Status: Terminated; Reported quantity: 19423 cases; Lot: Lot codes: 40179127 40179107 40179109 40179105 40179093 40179034 31279120 31279118 3057921 3057920 3057915 3057914 3057906 3057905 3047924 3047923. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- NxStage Medical, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to France, Netherlands, Spain, Sweden, and the UK.
- Lot
- Lot codes: 40179127 40179107 40179109 40179105 40179093 40179034 31279120 31279118 3057921 3057920 3057915 3057914 3057906 3057905 3047924 3047923
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALow Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1; 2)December 25, 2019
- FDANxStage PureFlow B Solution-RFP-402, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution isDecember 5, 2018
- FDANxStage PureFlow B Solution-RFP-400, Premixed Dialysate for Hemodialysis. NxStage PureFlow B Solution isDecember 5, 2018
- FDANxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5December 2, 2015
- FDANxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-301 Lactate 45, Potassium 1,July 2, 2014
- FDANxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-303 Lactate 45, Potassium 1,July 2, 2014
Questions about this recall
Was NxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-302 Lactate , 40, Potassium 1, Sodium 140, Calcium 3, Magnesium 1, Chloride 105, Glucose (mg/dL) 100, Batch Size (liters): 60 The SAK is one of the disposables used with NxStage PureFlow SL Dialysate Preparation System. The SAK is a disposable bag pre-filled with dialysate concentrate that is then filled with purified water prepared by NxStage PureFlow SL Dialysate Preparation System to prepare a batch of dialysate for use with the NxStage System One. SAKs are packaged in a case containing two SAKs. SAK product SKUs vary in the dialysate concentrate and volume of the final dialysate batch. recalled?
Yes. FDA published recall Z-1855-2014 on July 2, 2014. Reported hazard: SAK Dialysate Concentrate contain aluminum levels which exceed internal product specification
What should I do if I have this product?
Voluntary: Firm initiated · Letter