Hartmann Econolux Ref 416 105 Sterile Gauze sponge 4"x4" 12-ply 24 boxes/case 1200 total sponges For use in wound cleansing, debriding of wounds, wound packing, covering and swabbing
Published July 4, 2012 · Recall initiated: May 14, 2012
Summary
FDA records medical device Class II recall Z-1857-2012, published July 4, 2012 and initiated May 14, 2012. Recalling firm: Hartmann USA, Inc; Status: Terminated; Reported quantity: Lot 101546391 - 200 cs; Lot 101646398 - 200 cs; 101747392 - 120 cs; Lot 101850399 - 160 cs; Lot 101952390 - 240 cs; Lot 200106397 - 200 cs; Lot: Item # 416105 (Econolux): 101546391, 101646398, 101747392, 101850399, 101952309, 200106397. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Micro holes were detected in the laminate of the sterile barrier system of sterilized Hartmann Econolux and Hartmann Sterilux gauze products
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hartmann USA, Inc
- Distributor
- Nationwide distribution: USA including states of: AZ, CA, FL, GA, KY, MO, MS, NC, PA, SC, TN, TX, and VA.
- Lot
- Item # 416105 (Econolux): 101546391, 101646398, 101747392, 101850399, 101952309, 200106397
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHartmann Sterilux Ref 56910000 Sterile Premium gauze sponge 4"x4" 12-ply 24 boxes/case 1200 total sponges ForJuly 4, 2012
- FDAArrow Two-Lumen Central Venous Catheterization Set with Blue FlexTip(R) ARROWgard Blue(R) Catheter ProductJuly 4, 2012
- FDAFresenius GranuFlo (powder) Acid Concentrate Product Codes: OFD1201-3B, OFD1251-3B, OFD2123-3B, OFD2201-3B,July 4, 2012
- FDAGoBed+ . Model FL20E Beds are intended for medical purposes in both general patient care as well as criticalJuly 4, 2012
- FDAFresenius NaturaLyte Liquid Acid Concentrate Product Codes: 08-0231-4, 08-1001-0, 08-1201-8, 08-1231-3,July 4, 2012
- FDAGo Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as criticalJuly 4, 2012
Questions about this recall
Was Hartmann Econolux Ref 416 105 Sterile Gauze sponge 4"x4" 12-ply 24 boxes/case 1200 total sponges For use in wound cleansing, debriding of wounds, wound packing, covering and swabbing recalled?
Yes. FDA published recall Z-1857-2012 on July 4, 2012. Reported hazard: Micro holes were detected in the laminate of the sterile barrier system of sterilized Hartmann Econolux and Hartmann Sterilux gauze products
What should I do if I have this product?
Voluntary: Firm initiated · Letter