OEC Uroview 2800, Model number A349855. Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures.

Published July 4, 2012 · Recall initiated: November 20, 2006

Summary

FDA records medical device Class II recall Z-1865-2012, published July 4, 2012 and initiated November 20, 2006. Recalling firm: GE OEC Medical Systems, Inc; Status: Terminated; Reported quantity: 717 units; Lot: All Serial Numbers: P4-0004-L , P4-0005-L , P4-0006-R-RC1 , P4-0008-R , P4-0011-L , P6-0654-R , P4-0012-R , P4-0013-L , P4-0014-L , P4-0015-L , P4-0016-L , P4-0017-L , P4-0018-L , P4-0019-R , P4-0020-L , P4-0021-L , P4-0022-R , P4-0023-L , P4-0024-R , P4-0025-R , P4-0026-L , P4-0027-R , P4-0028-R , P4-0029-R , P4-0030-R , P4-0031-L , P4-0680-L , P4-0032-R , P4-0033-R , P4-0034-R , P4-0035-R , P4-0036-R , P4-0037-R , P4-0038-L , P4-0039-R , P4-0042-R , P4-0044-L , P4-0045-R , P4-0046-L , P4-0047-L-RC , P4-0048-L , P4-0049-R , P4-0050-L , P4-0051-L , P4-0052-R , P4-0053-L , P4-0054-R , P4-0074-R , P4-0055-L , P4-0058-R , P4-0059-L , P4-0062-L , P4-0064-L , P4-0067-L , P4-0075-L , P4-0068-R , P4-0070-L , P4-0071-L , P4-0073-R , P4-0076-R , P4-0078-R , P4-0081-L , P4-0082-R , P4-0388-L , P4-0084-R , P4-0085-R , P4-0087-R , P4-0088-L , P4-0089-L , P4-0090-L , P4-0092-R , P4-0093-R , P4-0094-R , P4-0095-R , P4-0097-R , P4-0098-R , P4-0099-L , P4-0100-R , P4-0101-R , P4-0102-R , P4-0103-R , P4-0104-R , P4-0105-R , P4-0107-L , P4-0108-R , P4-0110-L , P4-0111-L , P4-0112-L , P4-0113-R , P4-0114-L , P4-0661-L , P4-0115-R , P4-0117-R , P4-0118-R , P4-0119-L , P4-0120-L , P4-0121-L , P4-0123-L , P4-0124-R , P4-0127-L , P4-0128-R , P4-0131-R , P4-0132-R , P4-0133-L , P4-0138-R , P4-0140-L , P4-0141-R , P4-0142-R, P4-0154-R , P4-0696-L , P4-0143-L , P4-0146-R , P4-0147-L , P4-0149-R , P4-0211-R , P4-0151-L , P4-0153-R , P4-0155-R , P4-0159-L , P4-0160-L , P4-0161-L , P4-0162-R , P4-0163-R , P6-0529-L , P4-0164-R , P4-0165-L , P4-0167-R , P4-0168-R , P4-0169-R , P4-0170-R , P4-0172-R , P4-0175-L , P4-0176-L , P4-0177-R , P4-0178-R , P4-0179-L , P4-0181-L , P4-0183-R , P4-0185-R , P4-0301-R , P4-0187-L , P4-0188-L , P4-0190-R , P4-0191-L , P4-0194-L , P4-0195-R , P4-0196-L , P4-0197-L , P4-0199-R , P4-0201-R , P4-0202-R , P4-0203-L , P4-0204-R , P4-0205-R , P4-0206-L , P4-0209-L , P4-0212-L , P4-0215-L , P4-0216-R , P4-0223-R , P4-0232-L , P4-0237-R , P4-0238-L , P4-0239-R , P4-0240-L , P4-0241-L , P4-0242-R , P4-0244-R , P4-0246-R , P4-0247-R , P4-0248-R , P4-0451-L , P4-0254-L , P4-0255-R , P4-0256-R , P4-0257-L , P4-0258-L , P4-0259-L , P4-0445-L , P4-0260-R , P4-0263-L , P4-0264-R , P4-0268-R , P4-0659-R , P4-0269-R , P4-0271-R , P4-0273-L , P6-0715-L , P4-0274-R , P4-0277-L , P4-0279-R , P4-0685-L , P4-0280-R , P4-0282-L , P4-0281-L , P4-0284-L , P4-0285-L , P4-0288-R , P4-0289-L , P4-0292-L , P4-0293-R , P4-0294-L , P4-0296-R , P4-0297-R , P4-0298-R , P4-0299-L , P4-0300-R , P4-0318-L , P4-0302-L , P4-0304-R , P4-0305-L , P4-0306-R , P4-0308-L , P4-0309-R , P4-0310-L , P4-0313-L , P4-0315-L , P4-0316-R , P4-0457-R , P4-0319-R , P4-0321-L , P4-0322-L , P4-0325-R , P4-0326-R , P4-0327-L , P4-0328-R , P4-0329-L , P4-0332-L , P4-0333-R , P4-0335-R , P4-0339-L , P4-0340-R , P4-0341-R , P4-0342-L , P4-0343-L , P4-0345-R , P4-0346-L , P4-0347-L , P4-0348-R , P4-0349-L , P4-0353-L , P4-0354-R , P4-0355-L , P4-0356-L , P4-0359-R , P4-0361-R , P4-0362-R , P4-0365-R , P4-0367-R , P4-0370-L , P4-0373-R , P4-0377-L , P4-0380-L , P4-0381-L , P4-0382-R , P4-0385-R , P4-0386-R , P4-0389-L , P4-0391-L , P4-0393-R , P4-0394-R , P4-0401-R , P4-0402-R , P4-0405-L , P4-0406-L , P4-0407-R , P4-0408-R , P4-0409-R , P4-0410-R , P4-0411-L , P4-0413-L , P4-0416-R , P4-0415-L , P4-0417-R , P4-0418-L-RC , P4-0419-L , P4-0422-L , P4-0424-R , P4-0425-R , P4-0426-L , P4-0428-R , P4-0431-R , P4-0436-R , P4-0440-R , P4-0442-L , P6-0472-R , P4-0443-R ,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and "digital spot mode", and the system does not reboot.

What to do

FDA Mandated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE OEC Medical Systems, Inc
Distributor
Worldwide Distribution.
Lot
All Serial Numbers: P4-0004-L , P4-0005-L , P4-0006-R-RC1 , P4-0008-R , P4-0011-L , P6-0654-R , P4-0012-R , P4-0013-L , P4-0014-L , P4-0015-L , P4-0016-L , P4-0017-L , P4-0018-L , P4-0019-R , P4-0020-L , P4-0021-L , P4-0022-R , P4-0023-L , P4-0024-R , P4-0025-R , P4-0026-L , P4-0027-R , P4-0028-R , P4-0029-R , P4-0030-R , P4-0031-L , P4-0680-L , P4-0032-R , P4-0033-R , P4-0034-R , P4-0035-R , P4-0036-R , P4-0037-R , P4-0038-L , P4-0039-R , P4-0042-R , P4-0044-L , P4-0045-R , P4-0046-L , P4-0047-L-RC , P4-0048-L , P4-0049-R , P4-0050-L , P4-0051-L , P4-0052-R , P4-0053-L , P4-0054-R , P4-0074-R , P4-0055-L , P4-0058-R , P4-0059-L , P4-0062-L , P4-0064-L , P4-0067-L , P4-0075-L , P4-0068-R , P4-0070-L , P4-0071-L , P4-0073-R , P4-0076-R , P4-0078-R , P4-0081-L , P4-0082-R , P4-0388-L , P4-0084-R , P4-0085-R , P4-0087-R , P4-0088-L , P4-0089-L , P4-0090-L , P4-0092-R , P4-0093-R , P4-0094-R , P4-0095-R , P4-0097-R , P4-0098-R , P4-0099-L , P4-0100-R , P4-0101-R , P4-0102-R , P4-0103-R , P4-0104-R , P4-0105-R , P4-0107-L , P4-0108-R , P4-0110-L , P4-0111-L , P4-0112-L , P4-0113-R , P4-0114-L , P4-0661-L , P4-0115-R , P4-0117-R , P4-0118-R , P4-0119-L , P4-0120-L , P4-0121-L , P4-0123-L , P4-0124-R , P4-0127-L , P4-0128-R , P4-0131-R , P4-0132-R , P4-0133-L , P4-0138-R , P4-0140-L , P4-0141-R , P4-0142-R, P4-0154-R , P4-0696-L , P4-0143-L , P4-0146-R , P4-0147-L , P4-0149-R , P4-0211-R , P4-0151-L , P4-0153-R , P4-0155-R , P4-0159-L , P4-0160-L , P4-0161-L , P4-0162-R , P4-0163-R , P6-0529-L , P4-0164-R , P4-0165-L , P4-0167-R , P4-0168-R , P4-0169-R , P4-0170-R , P4-0172-R , P4-0175-L , P4-0176-L , P4-0177-R , P4-0178-R , P4-0179-L , P4-0181-L , P4-0183-R , P4-0185-R , P4-0301-R , P4-0187-L , P4-0188-L , P4-0190-R , P4-0191-L , P4-0194-L , P4-0195-R , P4-0196-L , P4-0197-L , P4-0199-R , P4-0201-R , P4-0202-R , P4-0203-L , P4-0204-R , P4-0205-R , P4-0206-L , P4-0209-L , P4-0212-L , P4-0215-L , P4-0216-R , P4-0223-R , P4-0232-L , P4-0237-R , P4-0238-L , P4-0239-R , P4-0240-L , P4-0241-L , P4-0242-R , P4-0244-R , P4-0246-R , P4-0247-R , P4-0248-R , P4-0451-L , P4-0254-L , P4-0255-R , P4-0256-R , P4-0257-L , P4-0258-L , P4-0259-L , P4-0445-L , P4-0260-R , P4-0263-L , P4-0264-R , P4-0268-R , P4-0659-R , P4-0269-R , P4-0271-R , P4-0273-L , P6-0715-L , P4-0274-R , P4-0277-L , P4-0279-R , P4-0685-L , P4-0280-R , P4-0282-L , P4-0281-L , P4-0284-L , P4-0285-L , P4-0288-R , P4-0289-L , P4-0292-L , P4-0293-R , P4-0294-L , P4-0296-R , P4-0297-R , P4-0298-R , P4-0299-L , P4-0300-R , P4-0318-L , P4-0302-L , P4-0304-R , P4-0305-L , P4-0306-R , P4-0308-L , P4-0309-R , P4-0310-L , P4-0313-L , P4-0315-L , P4-0316-R , P4-0457-R , P4-0319-R , P4-0321-L , P4-0322-L , P4-0325-R , P4-0326-R , P4-0327-L , P4-0328-R , P4-0329-L , P4-0332-L , P4-0333-R , P4-0335-R , P4-0339-L , P4-0340-R , P4-0341-R , P4-0342-L , P4-0343-L , P4-0345-R , P4-0346-L , P4-0347-L , P4-0348-R , P4-0349-L , P4-0353-L , P4-0354-R , P4-0355-L , P4-0356-L , P4-0359-R , P4-0361-R , P4-0362-R , P4-0365-R , P4-0367-R , P4-0370-L , P4-0373-R , P4-0377-L , P4-0380-L , P4-0381-L , P4-0382-R , P4-0385-R , P4-0386-R , P4-0389-L , P4-0391-L , P4-0393-R , P4-0394-R , P4-0401-R , P4-0402-R , P4-0405-L , P4-0406-L , P4-0407-R , P4-0408-R , P4-0409-R , P4-0410-R , P4-0411-L , P4-0413-L , P4-0416-R , P4-0415-L , P4-0417-R , P4-0418-L-RC , P4-0419-L , P4-0422-L , P4-0424-R , P4-0425-R , P4-0426-L , P4-0428-R , P4-0431-R , P4-0436-R , P4-0440-R , P4-0442-L , P6-0472-R , P4-0443-R ,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OEC Uroview 2800, Model number A349855. Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. recalled?

Yes. FDA published recall Z-1865-2012 on July 4, 2012. Reported hazard: Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and "digital spot mode", and the system does not reboot.

What should I do if I have this product?

FDA Mandated · Letter