Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10
Published July 3, 2019 · Recall initiated: April 2, 2019
Summary
FDA records medical device Class II recall Z-1868-2019, published July 3, 2019 and initiated April 2, 2019. Recalling firm: Thermo Fisher Scientific (Asheville) LLC; Status: Terminated; Reported quantity: 12 units; Lot: Serial Numbers: 300246811 300246812 300246814 300246815 300247403 300247404 300247406 300247407 300247409 300247789 300247791 300247792. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thermo Fisher Scientific (Asheville) LLC
- Distributor
- CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
- Lot
- Serial Numbers: 300246811 300246812 300246814 300246815 300247403 300247404 300247406 300247407 300247409 300247789 300247791 300247792
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMaquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229037A Model: VST66DFJuly 3, 2019
- FDAReplacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire,July 3, 2019
- FDATASE 400 Imaging systemsJuly 3, 2019
- FDALeadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA,July 3, 2019
- FDASmartSite Syringe Administration Set-Product Package Size 50, Catalog Number: 10798696July 3, 2019
- FDACO2 Incubators (General Purpose Incubator), model numbers 370, 371, 380, 381, 3307, 3308, 3310, 3311, 3540,April 17, 2013
Questions about this recall
Was Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10 recalled?
Yes. FDA published recall Z-1868-2019 on July 3, 2019. Reported hazard: Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit