FDAFDA · Food and drugs·Recall number Z-1877-2019

Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.

Published July 3, 2019 · Recall initiated: May 10, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Lot
product does not have lot or serial codes No GTIN/UDI
Risk classification
Class III
Area (FDA)
device
Country
United States