HEMASHIELD PLATINUM Woven Double Velour Vascular Graft-3 Branch, Product Code: M00202175912P0 - Product Usage: for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease.
Published May 13, 2020 · Recall initiated: April 7, 2020
Summary
FDA records medical device Class II recall Z-1876-2020, published May 13, 2020 and initiated April 7, 2020. Recalling firm: Maquet Cardiovascular, LLC; Status: Terminated; Reported quantity: N/A; Lot: Serial Numbers: 1174771828 1174771106 1174769941. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential small holes at the seam lines of Hemashield branches causing intra-operative bleeding at the seam lines of the branches of Hemashield grafts
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- Worldwide distribution - US Nationwide distribution and the countries of Dubai, Argentina, Austria, Belgium, Brazil, Japan, Canada, China, Germany, Spain, Finland, France, UK, Hongkong, Italy, Korea, Netherlands, Poland, Serbia, Sweden.
- Lot
- Serial Numbers: 1174771828 1174771106 1174769941
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,July 22, 2026
- FDAVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 EndoscopicOctober 1, 2025
- FDAAcrobat-i Positioner. Model Number C-XP-5000ZNovember 27, 2024
- FDAAcrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000ZNovember 27, 2024
- FDAHEMASHIELD PLATINUM Woven Double Velour Vascular Graft01 Branch, Product Code: M00202175834P0 - Product Usage:May 13, 2020
- FDAHEMASHIELD PLATINUM Woven Double Velour Vascular Graft-4 Branch, Product Code: M00202175930P0 - Product Usage:May 13, 2020
Questions about this recall
Was HEMASHIELD PLATINUM Woven Double Velour Vascular Graft-3 Branch, Product Code: M00202175912P0 - Product Usage: for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. recalled?
Yes. FDA published recall Z-1876-2020 on May 13, 2020. Reported hazard: Potential small holes at the seam lines of Hemashield branches causing intra-operative bleeding at the seam lines of the branches of Hemashield grafts
What should I do if I have this product?
Voluntary: Firm initiated · Letter