Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
Published June 12, 2024 · Recall initiated: May 1, 2024
Summary
FDA records medical device Class I recall Z-1877-2024, published June 12, 2024 and initiated May 1, 2024. Recalling firm: Getinge Usa Sales Inc; Status: Ongoing; Reported quantity: 120 units (US); Lot: Part Number 6682282; UDI/DI: 07325710000212: Serial Numbers: 17003 - 23454. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Getinge Usa Sales Inc
- Distributor
- Domestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.
- Lot
- Part Number 6682282; UDI/DI: 07325710000212: Serial Numbers: 17003 - 23454.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediatelyNovember 20, 2024
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediatelyNovember 20, 2024
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediatelyNovember 20, 2024
- FDAMEERA CL mobile operating table - Designed for the placement and positioning of the patient immediatelyNovember 20, 2024
- FDATortilla de Harina, 400 gr, 10 count, packaged in clear plastic bag, 20 packages per case.June 12, 2024
- FDAArrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REFJune 12, 2024
Questions about this recall
Was Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e. recalled?
Yes. FDA published recall Z-1877-2024 on June 12, 2024. Reported hazard: Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
What should I do if I have this product?
Voluntary: Firm initiated