FDAFDA · Food and drugs·Recall number Z-1880-2017

Omega Incline Platform Lift

Published May 3, 2017 · Recall initiated: April 12, 2017

Summary

FDA records medical device Class II recall Z-1880-2017, published May 3, 2017 and initiated April 12, 2017. Recalling firm: Savaria Concord Lifts, Inc.; Status: Terminated; Reported quantity: 74; Lot: s/n or lot number - Job Number 630046 through 630119. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Savaria Concord Lifts Inc has decided to initiate this recall of its Omega Incline Platform Lift because bolts can come loose causing the carriage to separate from the trolley and the unit to fail

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Savaria Concord Lifts, Inc.
Distributor
Worldwide Distribution - US Distribution and to North America.
Lot
s/n or lot number - Job Number 630046 through 630119
Risk classification
Class II
Area (FDA)
device
Country
Canada