Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T20VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: No
Published June 5, 2024 · Recall initiated: April 24, 2024
Summary
FDA records medical device Class II recall Z-1899-2024, published June 5, 2024 and initiated April 24, 2024. Recalling firm: Lucid Diagnostics, Inc.; Status: Ongoing; Reported quantity: 157 units (US); Lot: Lot Code: Product Sent to Clinical Site: Part #: EC20T20VI ; GTIN Number: 00860007510953 ; Lot #'s: 462 (Qty: 25), 486 (Qty: 32), 501 (Qty:100) Product at Distributor in Quarantine: Part #: EC20T20VI ; GTIN Number: 00860007510953 ; Lot #'s: 486 (Qty: 18), 540 (Qty: 97). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Lucid Diagnostics, Inc.
- Distributor
- US Nationwide distribution in the states of MN, MD.
- Lot
- Lot Code: Product Sent to Clinical Site: Part #: EC20T20VI ; GTIN Number: 00860007510953 ; Lot #'s: 462 (Qty: 25), 486 (Qty: 32), 501 (Qty:100) Product at Distributor in Quarantine: Part #: EC20T20VI ; GTIN Number: 00860007510953 ; Lot #'s: 486 (Qty: 18), 540 (Qty: 97)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VIJune 5, 2024
- FDAFDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinalJune 5, 2024
- FDAPhilips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software VersionJune 5, 2024
- FDAsmith&nephew OXINIUM Femoral Head 12/14 Taper, REF 71343204, 32 MM OD +4, Sterile; and smith&nephew, 12/14June 5, 2024
- FDAPhilips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version NumberJune 5, 2024
- FDATrevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XPJune 5, 2024
Questions about this recall
Was Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T20VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: No recalled?
Yes. FDA published recall Z-1899-2024 on June 5, 2024. Reported hazard: The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.
What should I do if I have this product?
Voluntary: Firm initiated · Letter