Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744
Published June 30, 2021 · Recall initiated: May 7, 2021
Summary
FDA records medical device Class II recall Z-1905-2021, published June 30, 2021 and initiated May 7, 2021. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 13,337 units; Lot: All Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- US: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia, West Virginia, Wisconsin, Wyoming, OUS: Andorra, Austria, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, Israel, Italy, Japan, Jordan, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, United Arab Emirates
- Lot
- All Lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAVentec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number:June 30, 2021
- FDA6" x 48" Elastic Poly Probe Cover, T Tip,Tele Fold, 2 bands Item ID: A-AU-0852NAJune 30, 2021
Questions about this recall
Was Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744 recalled?
Yes. FDA published recall Z-1905-2021 on June 30, 2021. Reported hazard: SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit