Back to search
FDAFDA · Food and drugs·Recall number Z-1905-2024

Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE, Catalog: 91412 TREVO NXT 4X28+TRAK21 2-PACK - CE, Catalog: 91413 TREVO NXT 4X41+TRAK21 2-PACK - CE, Catalog: 91414 TREVO NXT 6X37+TRAK21 2-PACK - CE, Catalog: 91415 TREVO NXT 3X32+PRO14 2-PACK - CE, Catalog: 91416 TREVO NXT 4X28+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914170 TREVO NXT 6X37+TRAK21+CAT6 3-PACK - CE, Catalog: SFP914190

Published June 5, 2024 · Recall initiated: April 23, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Neurovascular
Distributor
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
Lot
Catalog/UDI-DI/Lot: 90312/07613327299922/118881; 91412/07613327500684/48483279, 48483281, 48483294, 48696618, 48696619, 48696620, 48715713, 48834692, 48834695, 48834696; 91413/07613327500691/48483280, 48483284, 48483289, 48483291, 48483316, 48483317, 48696613, 48696622, 48696623, 48715711, 48715715; 91414/07613327500707/48447776, 48447778, 48447779, 48483290, 48483292, 48483293, 48483315, 48483318, 48483320, 48696624, 48715714, 48834690, 48834691, 48834694, 48834697, 48852726, 48852730, 48852731, 48852732, 48852734, 48852735, 48852737, 48852738; 91415/07613327500714/48483277, 48483283, 48483295, 48483296, 48483311, 48483312, 48696621, 48715712, 48715716, 48834688, 48834693, 48834698; 91416/07613327507904/47328863, 47328865, 47351815, 47351816, 47374814, 48500988, 48501006, 48501007, 48501008, 48501009, 48501010, 48501011, 48501012, 48501013, 48501014, 48540277, 48556780, 48556783, 48558375, 48588482, 48589876, 48641524, 50065509, 50065520, 50193572, 50260581, 50327385, 50327403, 50375186, 50375187, 50422577, 50422588, 50448274, 50738892, 50762691, 50762721, 50762722, 50786687, 50797491, 50797494, 50797495, 50886508, 51320490, 51320491, 51379394, 51379399, 51379400, 51379401, 51379402, 51379403, 51379404, 51379405, 51379406, 51379408, 51467927, 51489221, 51489230, 51489233, 51489234; SFP914170/07613327508055/48834689; SFP914190/07613327508079/48945914
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card