ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.
Published May 10, 2017 · Recall initiated: March 20, 2017
Summary
FDA records medical device Class II recall Z-1921-2017, published May 10, 2017 and initiated March 20, 2017. Recalling firm: Roho Inc.; Status: Terminated; Reported quantity: 5646 units; Lot: Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roho Inc.
- Distributor
- Nationwide, Sweden, Germany, New Zealand, Finland, Australia, Singapore, Korea, Luxembourg, Czech Republic, South Africa, Canada, Switzerland, United Arab Emirates, Spain, Austria, Hong Kong, Iceland, Japan, Canada, Italy, Ireland, France, United Kingdom
- Lot
- Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARENOVATION, THIN OSTEOTOME BLADE, 10 MM X 3 INCH, REF 71369410, STERILE R, QTY: (1)May 10, 2017
- FDAThe JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascularMay 10, 2017
- FDARENOVATION, THIN OSTEOTOME BLADE, 8 MM X 5 INCH, REF 71369408, STERILE R, QTY: (1)May 10, 2017
- FDASchiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is anMay 10, 2017
- FDARENOVATION, THIN OSTEOTOME BLADE, 10 MM X 3 INCH, REF 71369210, STERILE R, QTY: (1)May 10, 2017
- FDASecond generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part NumberMay 10, 2017
Questions about this recall
Was ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology. recalled?
Yes. FDA published recall Z-1921-2017 on May 10, 2017. Reported hazard: When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter