Back to search
FDAFDA · Food and drugs·Recall number Z-1921-2017

ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.

Published May 10, 2017 · Recall initiated: March 20, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Roho Inc.
Distributor
Nationwide, Sweden, Germany, New Zealand, Finland, Australia, Singapore, Korea, Luxembourg, Czech Republic, South Africa, Canada, Switzerland, United Arab Emirates, Spain, Austria, Hong Kong, Iceland, Japan, Canada, Italy, Ireland, France, United Kingdom
Lot
Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card