Philips Brilliance iCT SP Computed Tomography X-Ray System; Model # 728311
Published May 13, 2020 · Recall initiated: April 6, 2020
Summary
FDA records medical device Class II recall Z-1922-2020, published May 13, 2020 and initiated April 6, 2020. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 2; Lot: Serial # 200044 Additional unit as off 5/22/20: 200026. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America, LLC
- Distributor
- Worldwide distribution including one unit distributed to Iowa.
- Lot
- Serial # 200044 Additional unit as off 5/22/20: 200026
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for theApril 20, 2022
- FDAPhilips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A,November 4, 2020
- FDAPhilips SureSigns VM1SpO2, C02, Rec, model no. 863266 - Product Usage: The SureSigns VM1 is a patient monitorJune 17, 2020
- FDAPhilips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospitalMay 27, 2020
- FDAPhilips Brilliance iCT Computed Tomography X-Ray System; Model # 728306; UDI #s (01)00884838059474(21)100783,May 13, 2020
- FDAThe Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTINMay 13, 2020
Questions about this recall
Was Philips Brilliance iCT SP Computed Tomography X-Ray System; Model # 728311 recalled?
Yes. FDA published recall Z-1922-2020 on May 13, 2020. Reported hazard: The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
What should I do if I have this product?
Voluntary: Firm initiated