FDAFDA · Food and drugs·Recall number Z-1937-2024
MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010
Published June 12, 2024 · Recall initiated: April 22, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
- Lot
- UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States