MEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020
Published June 12, 2024 · Recall initiated: April 22, 2024
Summary
FDA records medical device Class II recall Z-1938-2024, published June 12, 2024 and initiated April 22, 2024. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Reported quantity: 5,075,234 syringes; Lot: UDI/DI 10080196306183, All lots in the format: 897YYMMXXXX. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
- Lot
- UDI/DI 10080196306183, All lots in the format: 897YYMMXXXX
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACURAD Alcohol Prep Pads, Sterile, Medium, 2-Ply, Contents: 5 boxes per Carton, 30 Boxes per Case, Single UseSeptember 3, 2025
- FDAReliOn, Sterile Alcohol Swabs, Skin Cleanser, 200 Swabs, Distributed by, Walmart, Inc., Bentonville, AR 72716September 3, 2025
- FDAGood Neighbor Pharmacy, Alcohol Prep Pads, Distributed By AmeriSource Bergen, 1 West First Avenue,September 3, 2025
- FDAMedline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004ASeptember 3, 2025
- FDAMEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010June 12, 2024
- FDAThe Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter ofJune 12, 2024
Questions about this recall
Was MEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020 recalled?
Yes. FDA published recall Z-1938-2024 on June 12, 2024. Reported hazard: It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
What should I do if I have this product?
Voluntary: Firm initiated · Letter